 The bold font indicates action items and regular text gives supplemental information. The bold font indicates action items and regular text gives supplemental information.
 +
 ====== Setting up Calibration GUI ====== ====== Setting up Calibration GUI ======
  
Line 20: Line 21:
  
   $>cd ~/daq/rce #move to rce directory   $>cd ~/daq/rce #move to rce directory
-  $>source ./scripts/setup-gen3.sh #setup environment. ALTERNATIVELY ANL can use source setupGUIandROOT.sh (if exists) to setup setup-gen3 and thisroot.sh+  $>source ./scripts/setup-gen3.sh #setup environment. ANL can use source setupGUIandROOT.sh to source setup-gen3 and thisroot.sh
   $>calibGUI #start the GUI to interface with the rce   $>calibGUI #start the GUI to interface with the rce
      
Line 29: Line 30:
  
 **On the GUI select Save As and select a location for the log file to be saved. I simply make a directory ($>mkdir ~/daq/rce/Logs) and choose this location. The Config Root Dir should be set and then left alone once everything has been initialized (I set it to /root/pixelrce/rce). The Data Dir should be set to /root/calibData by default which is preferred. Select a Run Number which by default will automatically increment, and set a Scan Type. For this demonstration choose THRESHOLD_SCAN. On the next tab "Config HalfStave A" make sure only as many front end windows as desired are selected as "Valid" and "Included". For this example there is only one included frontend, A1-1. Select Inlink and Outlink equal to 1 and rce equal to 0 (or whatever configuration is setup in /etc/hosts).  Last and most important, select the configuration file by clicking the CHOOSE button. The default file (94-22-04_25__cern__6999.cfg) can be found in moduleconfigs/fei4bexample/configs/. Be sure the Address parameter (see below) is set appropriately. Returning to the "Scan" tab you may click Start Run.** **On the GUI select Save As and select a location for the log file to be saved. I simply make a directory ($>mkdir ~/daq/rce/Logs) and choose this location. The Config Root Dir should be set and then left alone once everything has been initialized (I set it to /root/pixelrce/rce). The Data Dir should be set to /root/calibData by default which is preferred. Select a Run Number which by default will automatically increment, and set a Scan Type. For this demonstration choose THRESHOLD_SCAN. On the next tab "Config HalfStave A" make sure only as many front end windows as desired are selected as "Valid" and "Included". For this example there is only one included frontend, A1-1. Select Inlink and Outlink equal to 1 and rce equal to 0 (or whatever configuration is setup in /etc/hosts).  Last and most important, select the configuration file by clicking the CHOOSE button. The default file (94-22-04_25__cern__6999.cfg) can be found in moduleconfigs/fei4bexample/configs/. Be sure the Address parameter (see below) is set appropriately. Returning to the "Scan" tab you may click Start Run.**
 +
 +{{:itk:scans_calibgui.png?490|}}
 +{{ :itk:stave_calibgui.png?540|}}
  
 The default configuration file, (94-22-04_25__cern__6999.cfg) is not setup for the chip geometry currently at ANL-ASC (02-06-2016). To fix this open the file and edit the 'Address' parameter from 7 to 8. Save this altered file with a new name, (ex: default.cfg). The default configuration file, (94-22-04_25__cern__6999.cfg) is not setup for the chip geometry currently at ANL-ASC (02-06-2016). To fix this open the file and edit the 'Address' parameter from 7 to 8. Save this altered file with a new name, (ex: default.cfg).
Line 35: Line 39:
  
  
-**Once the test run is complete you will be able to go into the 'Plots' tab to examine the output of the scan. The results of the scan (including .root files and any calibrated configuration files) will be saved to the Data Dir (default ~/calibData) specified before the scan is started.**+**Once the test run is complete you will be able to go into the 'Plots' tab to examine the output of the scan. The results of the scan (including .root files and any calibrated configuration files) will be saved to the Data Dir (default is ~/calibData) specified before the scan is started.** 
 + 
 +{{:itk:plots_calibgui.png?500 |}}
  
 If for example, I ran a FDAC_TUNE there would be a folder in the calibData directory 'FDAC_TUNE_#####'. Inside this folder there would be histograms made from the scan, and more importantly in the config/fei4bexample/configs folder there will be a new .cfg file that can be loaded in at the start of a new scan that will use the calibration already performed by the previous FDAC_TUNE. Ideally one would run these calibrations one after the other and tune different parts of the chip. The GUI has a feature that lets this be done automatically (see next section). If for example, I ran a FDAC_TUNE there would be a folder in the calibData directory 'FDAC_TUNE_#####'. Inside this folder there would be histograms made from the scan, and more importantly in the config/fei4bexample/configs folder there will be a new .cfg file that can be loaded in at the start of a new scan that will use the calibration already performed by the previous FDAC_TUNE. Ideally one would run these calibrations one after the other and tune different parts of the chip. The GUI has a feature that lets this be done automatically (see next section).
Line 74: Line 80:
 The logic behind the primlist file is fairly obvious and can be manipulated and altered as the user desires different parameters, orderings, or scans.  The logic behind the primlist file is fairly obvious and can be manipulated and altered as the user desires different parameters, orderings, or scans. 
  
-**It is most instructive to run a diagnostic scan such as the THRESHOLD_SCAN on the default .cfg file, then again on the last .cfg file after the primlist has run (for example in the ~/calibData/NOISESCAN_####/config/fei4bexample/configs/ directory where #### is the last scan number). Use ROOT to compare the histograms in the two THRESHOLD_SCANs before and after to ensure everything was executed as expected.**+**It is most instructive to run a diagnostic scan such as the THRESHOLD_SCAN on the default .cfg file, then again on the last .cfg file after the primlist has run (for example in the ~/calibData/NOISESCAN_####/config/fei4bexample/configs/ directory where #### is the last scan number). Use ROOT to compare the histograms in the two THRESHOLD_SCANs before and after to ensure everything was executed as expected. See below for example.** 
 + 
 +{{:itk:threshold_before_fit.png?575 |}} 
 +{{ :itk:threshold_after.png?375|}} 
 + 
 +  
 +  